Abstract:Objective To evaluate the influence factor on pathogen detection of blood stream infection by BD BACTEC 9240 automated blood culture system.Methods The results of BD BACTEC 9240 automated blood culture system was studied retrospectively,and the pathogen detection rate in double sets of blood culture was analyzed.The influences of antibiotic therapeutic concentration and delivery time were evaluated in simulated blood cultures in vitro.Results Double sets of blood culture can increase the detection rate of bloodstream infection.Escherichia coli and Staphylococcus aureus could be detected in therapeutic concentrations of gentamicin,cefotaxime,vancomycin,piperacillin/tazobactam and levofloxacin in simulated blood cultures in vitro,but be not detected by imipenem therapeutic concentration.Streptococcus pneumoniae was inhibited by the selected antibiotic therapeutic concentration.Delivery time would not affect the pathogen detection,but decreased the time to detection(TTD) in blood cultures.Conclusions Double sets of blood culture can increase the detection rate of bloodstream infection.Antibiotics would inhibit some pathogen growth in blood cultures,leading false negative.Delivery time would not affect pathogen detection,but would change TTD.TTD is effected by many factors,could not be a surrogate for the initial bacterial concentration in bloodstream infection.