Microwave-assisted digestion of glass samples for the determination of cadmium,arsenic and lead by atomic fluorescence spectrometry

XIN Juan-juan,HE Rong-li,YANG Qiu-ju,WU Li,HOU Xian-deng
DOI: https://doi.org/10.3969/j.issn.1004-1656.2006.04.027
2006-01-01
Abstract:Closed-vessel microwave-assisted digestion was utilized for the decomposition of hard-to-dissolve optical glass samples for the determination of the ultra trace amount of cadmium by electrothermal evaporation atomic fluorescence spectrometry(ETV-AFS)and arsenic and lead by hydride generation atomic fluorescence spectrometry(HG-AFS).The limits of detection were found to be 0.08,0.1 and 0.7ngmL~(-1)for cadmium,arsenic and lead,respectively.Relative standard deviation(RSD)was between 2%-5% which was adequate for trace analysis.The analytical results were in good agreement with those obtained by inductively coupled plasma mass spectroscopy(ICP-MS)for cadmium and inductively coupled plasma atomic emission spectrometry(ICP-AES)for arsenic and lead.
What problem does this paper attempt to address?