Electrophoretic chip based on special wettability surfaces for detection of heavy metals
Shuyue Jiang,Haifeng Zhang,Panpan Wang,Zhaoxin Li
DOI: https://doi.org/10.1016/j.microc.2023.109432
IF: 5.304
2023-09-30
Microchemical Journal
Abstract:Electrophoretic chip coupling to electrochemical detection enables the separation, monitoring, and on-site testing of multiple analytes simultaneously. However, bonding technology and detection sensitivity of the electrophoretic chip have severely diminished its value in industrial production. In this paper, a novel open PDMS/PCB electrophoretic chip based on special wettability surfaces was proposed. The developed chip employs the combination of superhydrophobic surface and super-hydrophilic microchannels to complete the function of bonding technology, avoiding the unreliability of the bonding process. The developed chip applied the super-hydrophilic porous MXene/CeO 2 /PVA hydrogels electrode, enhancing the sensitivity of chip. Finally, the developed chip was employed to simultaneously detect Pb 2+ , Hg 2+ , Cu 2+ , and Cd 2+ in laboratory and tap water via differential pulse voltammetry (DPV). It exhibited excellent performance with a low limit of detection (Pb 2+ 0.96 nM, Hg 2+ 0.87 nM, Cu 2+ 0.80 nM, and Cd 2+ 0.97 nM). The features of cost-effectiveness, high-reliability, and high-sensitivity make this electrophoretic chip useful for on-site monitoring of multiple analytes in industrial and medical fields.
chemistry, analytical