Evaluation of Malignant or Benign Bile Duct Strictures by Intraductal US

LIU Feng,JIN Zhen-dong,LI Zhao-shen,ZHANG Jing,ZOU Duo-wu,FANG Lin,YANG Li,MAN Xiao-hua
2007-01-01
Abstract:Objective: The aim of this study was to determine whether intraductal US is of assistance in distinguishing benign from malignant biliary strictures. Methods: A retrospective review was undertaken of 57 patients with bile duct strictures who underwent ERCP and IDUS from January 2003 to June 2006.A 20 MHz over-the-guidewire intraductal US catheter probe was used during ERCP for further examination of the strictures. Results: Final diagnoses of malignant strictures (31 patients) were confirmed histopathologically;confirmation of benign stricture (26 patients) was based on negative tissue sampling plus extended clinical follow-up.Sensitivity,specificity, positive predictive values, negative predictive values and accuracy of ERCP cytology brushing for malignant bile duct strictures were 39.1%,100%, 100%,48.1% and 61%. Sensitivity,specificity,positive predictive values,negative predictive values and accuracy of IDUS for malignant bile duct strictures were 80.6%,80.8%, 83.3%, 77.8% and 80.7%. Conclusion:Intraductal US is safe and can improve on the ability at ERCP so as to distinguish benign from malignant biliary strictures.