Test Methods and Instruments for Rheological Performance of Magnetorheological Fluids

Changrong Liao,Aijun Chen,Honghui ZHANG,Miao Yu,Weimin Chen
DOI: https://doi.org/10.3321/j.issn:0254-3087.2008.12.003
2004-01-01
Abstract:On the basis of technical requirements in the development of magnnetorheological(MR) fluid devices and in the evaluation of MR fluid materials,several constitutive models of MR fluid rheological properties are comprehensively analyzed,in which MR fluid constitutive models are expressed as a function of shear stress versus shear rate under the conditions of both special magnetic induction and special temperature.A test methodology based on double rotational coaxial cylinder shear mode is presented to realize high shear rate and high magnetic field.Based on the differential equation and its rational boundary conditions from analyzing stress equilibrium of MR fluid micro-unit,the theoretical relationship between nominal shear stress and transmission torques is analytically established,the theoretical relationship between nominal shear strain rate and rotor angular velocity is also comprehensively established,and corresponding algorithms are worked out.The relationship between average value of magnetic indication in the annular channels and excitation electrical currents are experimentally obtained.A rheometer is fabricated,in which the torque sensor,angular velocity sensor,infrared temperature sensor and ampere meter are reasonably selected.The rheometer was used to test the rheological parameters of certain MR fluid and the test results are in good agreement with those obtained by Department of Engineering,Chongqing University.