The Design of EEG Signal Acquisition Pre-processing Circuit

陈真诚,钟靖
2009-01-01
Abstract:Objective: EEG signal acquisition pre-processing circuit is designed in order to extract the weak EEG under much noise. Methods: The voltage of EEG is very low, usually between 5 μV ~100 μV, so a variable gain circuit is designed , which enable to magnify the EEG signals in 10000~50000 times. Meanwhile, the EEG signals are under the interference of the polarizing voltage of electrodes, voltage of high frequency, voltage of 50 Hz industrial frequency, so EEG preamplifier circuit, EEG high -pass filter, EEG low-pass filter and 50Hz-trap filter are designed in order to get rid of interference. A signal isolation circuit is designed in consideration of safety. Especially the preamp circuit is the key to the entire circuit. Results: After treatment of each module, it obtained high-quality EEG signal in order to follow-up application. Conclusions: The results from the experiment indicate that the design can extract the weak EEG signals and is of good practical use.
What problem does this paper attempt to address?