Chemiluminescence Flow Injection Analysis of Menadione Sodium Bisulfite Based on Luminol Reaction

Yuming Huang,Zhujun Zhang,Daojian Zhang
DOI: https://doi.org/10.1080/00032710008543215
2008-01-01
Analytical Letters
Abstract:A chemiluminescence(CL) flow system is described for the determination of menadione sodium bisulfite based on its repression on the chemiluminescence(CL) emission produced upon mixing hexacyanoferrate(III) solution with an alkaline luminol solution in the absence of co-oxidizer. The system responds linearly to menadione sodium bisulfite concentration in the range 0-1 mu g/mL with a detection limit (3 sigma) of 0.01 mu g/mL. Relative standard deviation (RSD) of 0.16% for 0.4 mu g/mL menadione sodium bisulfite (n=11). The system has been successfully applied to the determination of menadione sodium bisulfite in tablets and injections.