PS220. Thermo-Mechanical Resistance of a Nanocomposite Polymer Exposed to Simulated in Vivo Hydrodynamic Fatigue for Ten Years in Development of a Sutureless Endovascular Stent Graft

Mital Y. Desai,Raheleh Bakhshi,Arnold Darbyshire,Max Ahmed,James Eaton-Evans,Xiang Zhou,Zhong You,Alexander M. Seifalian,George Hamilton
DOI: https://doi.org/10.1016/j.jvs.2011.03.167
IF: 4.86
2011-01-01
Journal of Vascular Surgery
Abstract:To physiologically test thermo-mechanical resistance of a nanocomposite polymer based on polyhedral oligomeric silsesquioxane (POSS) and poly (carbonate-urea) urethane (PCU) to assess its feasibility as graft material for a sutureless aortic stent-graft by exposing it to simulated in vivo hydrodynamic fatigue according to FDA guidelines. Aortic stent-grafts (n=4) were tested in 37 degree Celsius distilled water using in vitro accelerated circumferential pulsatile model. After analysis for 400-million cycles equivalent to 10-years in human body, the mechanical and elastic properties were determined by Fourier transform infrared spectroscopy (FTIR), radial stress-strain studies, and linear elasticity tests. Dynamic scanning calorimetry (DSC) and Thermo-mechanical analysis (TMA) were used to assess thermal resistance. Comparison was made with zero-cycled control. All stent-grafts (n=4; diameter 25±0.1 mm; length 25.3±0.5 cm) successfully completed accelerated pulsatile fatigue testing at 37°C and pulse pressure 94±14 mm Hg over 12 months. FTIR showed increased intensity of -NHCO- bonds (62% transmittance in tested specimen compared to 47% in control) but there was no difference in intensity of other major bonds and no significant sign of degradation. Tensile strain of fatigue-tested polymer compared favourably with zero-cycled control at 50% to 500% stress (P=0.69). When stretched up to 500% original length, there was no difference in linear elasticity (deformation 13.3% in tested specimen compared to 12.3% in control; P=0.74). DSC and TMA showed decrease in glass transition temperature up to 4°C but thermo-tropic transition was comparable to control. Simulated physiological in vivo hydrodynamic fatigue has no significant degradative effect on an innovative stent-graft from POSS-PCU nanocomposite polymer. Sutureless technology incorporating nitinol stents proved to be robust with no separation over accelerated 10-year cycle.
What problem does this paper attempt to address?