Determination of Granisetron -Sufentanil and Granisetron -Fentanyl -Tramadol Admixtures in Patient-Controlled Analgesia Pumps by HPLC

KUANG Yu-shan,YIN Yan,YANG Ying-ying,ZHANG Wen-sheng,LIU Jin
2009-01-01
Abstract:OBJECTIVE To establish a method to determine the contents of Granisetron -Sufentanil and Granisetron -Fentanyl -Tramadol admixtures in Patient-controlled analgesia pumps. METHODS HPLC method was adopted. An Agilent hypersil C18 column (125 mm ×4 mm,5 μm) was used. The mobile phase consisted of 0.02 mol·L-1 potassium dihydrogen phosphate solution(0. 6% triethylamine, using phosphoric acid to adjust pH4) -acetonitrile. The flow rate was 1 mL·min~(-1) and the detection wavelength was 220 nm. RESULTS The linear range of Granisetron was 5 -50 μg·mL~(-1) (r = 0. 9999) and the average recovery was 99. 97 %. The linear range of Sufentanil was 0.25 -4.00 μg·mL~(-1) (r =0.9994) and the average recovery was 98.75%. The linear range of Tramadol was 20 -70 μg·mL~(-1) (r = 0.9996) and the average recovery was 99.80%. The linear range Of Fentanyl was 2.5 -20. 0 μg·mL~(-1) (r = 0. 9999)and the average recovery was 100.53%. CONCLUSION The method is rapid, simple, accurate, and can be used for the determination of Granisetron -Sufentanil and Granisetron -Fentanyl -Tramadol admixtures in Patient -controlled analgesia pumps.