Effect of Scan FOV on CT Value and Material Decomposition Measurement in Dual Energy CT: a Phantom Study

赵永为,郭小超,王霄英,刘建新
DOI: https://doi.org/10.3969/j.issn.1000-0313.2012.03.012
2012-01-01
Abstract:Objective:To study whether change of scan field of view(FOV) has effect on measurement of CT value and material decomposition when using dual energy CT scan(Gemstone Spectral Imaging,GSI).Methods:Seventeen tubes of water and iodine solution with different concentrations of iodine contrast were scanned with GSI mode.The iodine concentrations of these tubes were 0(water only),0.1,0.2,0.3,0.4,0.5,0.6,0.8,1.0,2.0,3.0,4.0,5.0,6.0,8.0,10.0 and 20.0mgI/ml,respectively.All tubes were scanned twice with large and medium FOV,respectively(Group A:Large FOV;Group B:Medium FOV).The CT value was measured on CT images with mixed energy.Aw4.4 workstation was used to calculate water and iodine decomposition in each tube.Results:The calculated iodine concentration was highly consistent with actual iodine concentration in the tubes(rA=0.9995,rB=0.9996,P=0.000).Measured CT value and calculated water concentration were statistically different between the two FOV groups(P0.01).There was no statistical difference on calculated iodine concentrations between the two FOV groups(P=0.19).Conclusion:Dual energy CT images can reflect the content of iodine in different concentrations and can be used for quantitative analysis.Changing scan FOV will affect the measurement of CT value,but has no effect on calculation of iodine concentration.