Separation and Detection of Iohexol with Cation-Exchange Column

陈斌,李丹,李曙光,马晓迅,陈国亮,李蓉
DOI: https://doi.org/10.16155/j.0254-1793.2012.10.028
2012-01-01
Abstract:Objective:To establish an HPLC method for the separation and detection of iohexol.Methods:The separation of iohexol and impurities was performed on a strong acid cation ion exchange column(Ultimate XB-SCX);mixture of acetonitrile and water(90∶ 10) was used as mobile phase with a flow rate of 2.0 mL·min-1,the injection volume was 10 μL.Results:The calibration curve of iohexol was linear in the range of 0.2-4.0 mg·mL-1 and the correlation coefficient(r) was 0.9994.The recoveries of the method were between 98% and 102%,and RSDs were all below 0.5%(n=5).The minimum limit of detection(LOD) was 0.5 μg·mL-1.Conclusion:The established method has a low detection limit,good accuracy and reproducibility.The study provides a fast and effective way for iohexol purification and detection.