System Errors Calibration for GLS-1500 Scanner

GAO Bao-hua,JIANG Li-xing,ZOU Xue-bing,ZHANG Tao
DOI: https://doi.org/10.3969/j.issn.1672-5867.2013.05.017
2013-01-01
Abstract:An in-depth study on the system error sources of 3D laser scanner is presented,and the three-stage method,horizontal sequence mark method and vertical sequence mark method are designed and applied to calibrating and correcting the ranging additive constant,horizontal angle error and vertical angle error of GLS-1500 respectively,and the effect of the mark center extraction accuracy on the calibration result is analyzed in the angle calibration experiments.