Design, Testing and Control of a Magnetorheological Damper for Knee Prostheses
Hounan Song,Yu Cao,Wei Chen,Lei Ren,Yongxin Ma,Kunyang Wang,Xu Wang,Yao Zhang,Luquan Ren
DOI: https://doi.org/10.1007/s42235-024-00535-1
IF: 2.995
2024-06-01
Journal of Bionic Engineering
Abstract:This study aims to develop a magnetorheological (MR) damper for semi-active knee prostheses to restore the walking ability of transfemoral amputees. The core dimensions of the MR damper were determined via theoretical magnetic field calculations, and the theoretical relationship between current and joint torque was derived through electromagnetic simulation. Then, a physical prototype of the semi-active prosthetic knee equipped with the MR damper was manufactured. Based on the data obtained from angle sensor, pressure sensor (loadcell), and inertial measurement unit (IMU) on the prosthesis, a matching control algorithm is developed. The joint torque of the MR damper can be adaptively adjusted according to the walking speed of the amputee, allowing the amputee to realize a natural gait. The effectiveness of the control program was verified by the ADAMS and MATLAB co-simulation. The results of the test and simulation show that the MR damper can provide sufficient torque needed for normal human activities.
robotics,engineering, multidisciplinary,materials science, biomaterials