Simultaneous determination of benzene series,ester and ketone in air by chromatography with dissolvent desorption
Hu Mei-hua,Zhu Hui-fang,Li Liang,Jiang Yong
DOI: https://doi.org/10.3969/j.issn.1004-8685.2008.06.009
2008-01-01
Abstract:Objective:A method for the simultaneous determination of benzene series,ester and ketone in air by gas chromatography was developed.Methods:Benzene series,ester and ketone in air were collected with charcoal tube,then were injected after being desorbed with carbon disulfide through kf(Poly(dimethylsiloxane)) capillary column(50 m×0.32 mm×1.0 μm).The primary temperature was set at 50℃ remaining 4 minutes,and the last temperature was 120℃ remaining 1 minute with elevate temperature ratio 5℃ per minute.Apart-flow rate was splitless.Results:Retention time for acetone,ethyl aceate,benzene,4-methyl-2-pentanone,toluene,butyl acetate,m-xylene and o-xylene were 3.55,5.69,7.13,9.54,10.88,12.47,15.11 and 16.07 min,respectively.The linear ranges were 1.1~440 μg/ml,1.1~442.8 μg/ml,0.5~200 μg/ml,1.2~490.4 μg/ml,0.5~200 μg/ml,1.1~433.2 μg/ml,0.5~200 μg/ml and 0.5~200 μg/ml.The limits of detection were 1.1,1.1,0.5,1.2,0.5,1.1,0.5 and 0.5 μg/ml.The coefficient of correlation was 0.9988, 0.9982,0.9996,0.9995,0.9997,0.9995,0.9997 and 0.9997.Conclusion:The method is applicable in simultaneously determining benzene series, ester and ketone in air.