Determination of Pralidoxime Iodide in Human Plasma and Urine by Reversed - Phase High Performance Liquid Chromatography

金鸣,高娃,周亦武,陈玉明,杨德隆
DOI: https://doi.org/10.3969/j.issn.1001-5728.2006.02.008
2006-01-01
Journal of Forensic Medicine
Abstract:Objective To develop a reversed-phase high performance liquid chromatography(HPLC) method for the purpose of more accurate and rapid analysis of pralidoxime iodide in human plasma and urine.Methods Standard pralidoxime iodide was added to the blank plasma and urine after the samples were pretreated.The method for pretreating samples,the linear range,the precision and the recoveries in the plasma were investigated by using human plasma and urine spiked with standard pralidoxime iodide.The HPLC method has been used to determine the concentration of pralidoxime iodide in some practical cases suspicious of organophosphorous intoxication.Results The linear range were 0.5~8.0 μg/ml with r=0.9988 in plasma and r=0.9968 in urine.The limit of detection was 0.5μg/ml.The intra-and inter-day precisions of assay for pralidoxime iodide were less than 3.5% and 4.4%((n=5)) in plasma,and 3.8% and 5.3% in urine,respectively.The recoveries of pralidoxime iodide were more than 96.7±2.9% in plasma and 93.7±3.8% in urine.Conclusion The HPLC method developed for quantitative analysis of pralidoxime iodide is simple,rapid and accurate,which is suitable for monitoring the concentration of pralidoxime iodide in forensic and clinical practice.
What problem does this paper attempt to address?