Microarray H2o2 Electrodes As Base Elements Of High-Response Glucose Sensors

Jianzhong Zhu,Jiali Wu,Chengyun Tian,Wei Wu,Hongwei Zhang,Deren Lu,Guoxiang Zhang
DOI: https://doi.org/10.1016/0925-4005(93)01166-2
1994-01-01
Abstract:A high-response glucose sensor has recently been developed by modifying an immobilized GOD (glucose oxidase) membrane and coating Nafion diffusion-limiting membrane over an amperometric H2O2 MAE (microarray electrode), which is fabricated by using standard planar processing. The chip of the sensor is 2.5 mm by 15 mm. The MAE is composed of an Ag/AgCl RE (reference electrode), at Pt CE (counter electrode) and two sets of Pt WEs (working electrodes), each set consisting of five bands. Silicon nitride is chosen as the top insulator. Six MAEs different in both width and length of the band are examined and the MAE with the narrowest bandwidth (10 mum) exhibits the best electrochemical performance. The glucose sensor based on the MAE B1-2 presents a linear range of 0.5-40 mM glucose at 37-degrees-C with a 1115 M phosphate buffer solution (pH = 7.0) as supporting electrolyte. The sensitivity, background current, response time, coefficient of variation and long-term stability of the sensor are 7.1 +/- 0.5 nA/mM, 2 +/- 0.5 nA, 30 s, 3.4% and 6 days (or 250 assays), respectively.