The in vitro study on transfer of bacteria DNA across dialyzers with different dialysis membranes

CHEN Huan,ZHAO Ban,YAN Yu,WANG Mei
DOI: https://doi.org/10.3969/j.issn.1671-4091.2013.05.005
2013-01-01
Abstract:Objective To assess the transfer of bacteria DNA across the dialyzers with cellulose triacetate (CTA, low-flux) and polysulfone (low-flux) membranes. Methods Two dialyzers equipped with CTA (low-flux, 130G) membrane or with polysulfone (low-flux, F6hps) membrane were used in this study. The experiment was carried out in saline-saline model. The dialysate was mixed with E. coli at 4 levels (2.06 x 10 2 , 1.603 x 10 3 , 2.671 x 10 3 , 1.336x 10 4 cells/ml), and the in vitro dialysis was performed under diffusive condition for 4 hours. Ultraviolet spectrophotometry and LAL test were used to detect bacteria DNA and endotoxin, respectively, in blood side samples. Bacteria DNA concentrations in blood side sample and dialysate were compared between the two kinds of dialyzers. Results After the dialysis for 4 hours, LAL in blood side samples from the two kinds of dialyzers remained negative, but E. coli DNA increased along with the increase of E. coli concentration in dialysate. When dialysate contained E. coli of 1.603 × 10 3 ,2.671 ×10 3 , 1.336 × 10 4 cells/ml, bacterial DNA in blood side samples from the dialyser with CTA membrane were significantly higher than those from the dialyser with polysulfone membrane (P<0.05). Conclusion Under the identical conditions, both the two dialyzers have the permeability for bacteria DNA but not for endotoxin. CTA low-flux dialyzer has higher permeability for bacteria DNA than polysulfone low-flux dialyzer.