Modification of Medical Metals by Ion Implantation of Copper

Y. Z. Wan,G. Y. Xiong,H. Liang,S. Raman,F. He,Y. Huang
DOI: https://doi.org/10.1016/j.apsusc.2007.06.031
IF: 6.7
2007-01-01
Applied Surface Science
Abstract:The effect of copper ion implantation on the antibacterial activity, wear performance and corrosion resistance of medical metals including 317L of stainless steels, pure titanium, and Ti–Al–Nb alloy was studied in this work. The specimens were implanted with copper ions using a MEVVA source ion implanter with ion doses ranging from 0.5×1017 to 4×1017ions/cm2 at an energy of 80keV. The antibacterial effect, wear rate, and inflexion potential were measured as a function of ion dose. The results obtained indicate that copper ion implantation improves the antibacterial effect and wear behaviour for all the three medical materials studied. However, corrosion resistance decreases after ion implantation of copper. Experimental results indicate that the antibacterial property and corrosion resistance should be balanced for medical titanium materials. The marked deteriorated corrosion resistance of 317L suggests that copper implantation may not be an effective method of improving its antibacterial activity.