A Novel Thin-Layer Amperometric Detector Based on Chemically Modified Ring-Disc Electrode and Its Application for Simultaneous Measurements of Nitric Oxide and Nitrite in Rat Brain Combined with in Vivo Microdialysis

LQ Mao,GY Shi,Y Tian,HY Liu,LT Jin,K Yamamoto,SG Tao,JY Jin
DOI: https://doi.org/10.1016/s0039-9140(98)00027-7
IF: 6.1
1998-01-01
Talanta
Abstract:A novel thin-layer amperometric detector (TLAD) based on chemically modified ring-disc electrode and its application for simultaneous measurements of nitric oxide (NO) and nitrite (NO(2)(-)) in rat brain were demonstrated in this work. The ring-disc electrode was simultaneously sensitive to nitric oxide (NO) and nitrite (NO(2)(-)) by modifying its inner disc with electropolymerized film of cobalt(II) tetraaminophthalocyanine (polyCoTAPc)/Nafion and its outer ring with poly(vinylpyridine) (PVP), respectively. The ring-disc electrode was used to constitute a novel TLAD in radial flow cell for simultaneous measurements of NO and NO(2)(-) in rat brain combined with techniques of high performance liquid chromatography (HPLC) and in vivo microdialysis. It was found that the basal concentration of NO in the caudate nucleus of rat brain is lower than 1.0x10(-7) mol l(-1), NO(2)(-) concentration is 5.0x10(-7) mol l(-1) and NO exists in brain maybe mainly in the form of its decomposed product.
What problem does this paper attempt to address?