Determination of Dopamine in Rat Striatum by Microdialysis and High-Performance Liquid Chromatography with Electrochemical Detection on a Functionalized Multi-Wall Carbon Nanotube Electrode.

Li Lin,Peihong Qiu,Lizhu Yang,Xuni Cao,Litong Jin
DOI: https://doi.org/10.1007/s00216-005-0275-3
2006-01-01
Abstract:High performance liquid chromatography coupled with microdialysis sampling and electrochemical detection (HPLC–ECD) has been used to determine dopamine (DA). In the HPLC–ECD a multi-wall carbon nanotube electrode chemically modified with carboxyl groups (MWNT-COOH CME) was used as the working electrode for determination of DA. The results indicated that the MWNT-COOH CME enabled efficient electrocatalytic oxidation of DA with relatively high sensitivity and stability and long life. Peak currents for dopamine were linearly dependent on concentration in the range 5.0×10−9 to 5.0×10−5 mol L−1 and the calculated detection limit (S/N=3) was 2.5×10−9 mol L−1. The method had been successfully used to measure dopamine in rat striatal microdialysate. To study the physiological effect of nitric oxide (NO) on striatal release of DA, 0.5 mmol L−1 sodium nitroprusside (SNP) was a continuously perfused into rat striatum. This resulted in a 46% increase in DA basal level.
What problem does this paper attempt to address?