Development of a Miniaturized Electrochemiluminescent Detection Cell and Its Performance

Bing-zheng TIAN,Wen-ying GUO,Yi-feng TU
DOI: https://doi.org/10.3969/j.issn.1008-2794.2009.04.014
2009-01-01
Abstract:In this paper,a novel miniaturized electrochemiluminescent(μ-ECL) cell used for flow injection analysis was designed and fabricated.Main effective factors on the cell performance were optimized.The ECL intensity of luminol in this cell increased linearly with the concentration in the range of 5×10-10 to 5×10-8mol/L with the correlation coefficient of 0.9998 and the detection limit of 1×10-10mol/L.Furthermore,a detection limit of 5×10-5mol/L was achieved for H_2O_2 flow injection analysis,and the linear range is from 1×10-4 to 5×10-3mol/L with the correlation coefficient of 0.9993.Based on this foundation,the application of the cell to real samples was studied on anti-oxygen ability evaluation of fruits.The results indicate that,it is clear that the anti-oxygen ability of red grape seed was stronger than the flesh,for 121 and 5.1mgH_2O_2/g respectively.The cell has good reproducibility in real sample determination.It has the advantages of high sensitivity,the practicability of flow injection analysis and simple and fast operation.
What problem does this paper attempt to address?