Preparation of a polyaniline-o-aminophenol polymer film modified electrode and its application

FAN Li-fang,FAN Yue-qin,MENG Shuang-ming,YANG Yang,ZHAI Hong,GUO Yong
2013-01-01
Abstract:A glassy carbon electrode(GCE) modified by electrochemical oxidation of aniline(AN) and o-aminophenol(OAP) has been prepared by cyclic voltammetry(CV) in this article.The polymer film modified electrode was applied to the determination of AA.To obtain the optimal results,some experimental parameters have been discussed,such as the concentration ratio of OAP and AN,pH of PBS buffer solution.The result showed that the PAN-OAP/GCE obtained at the concentration ratio of OAP to AN(1:14) and pH 6.80,can maintain high electroactivity and stability.Under the optimal experiment conditions,differential pulse voltammetry(DPV) has also been employed to determine AA.The result indicated that the PAN-OAP/GCE polymer film modified electrode in PBS(pH 6.80) buffer solution showed good electrocatalytic activity towards the oxidation of AA.Besides,the peak current of AA on PAN-OAP/GCE was linearly related to its concentration in the range of 1.50×10-8 -2.12×10-6 mol/L with a detection limit of 5.0×10-9 mol/L.The equation of linear regression is ip=1.0344c+0.0183 with a correlation coefficient of 0.9988.The film modified electrodes prepared by the electrochemical method in the paper has high sensitivity and selectivity,which can be used to determine AA in real samples,human blood.The recoveries of AA from the real samples,human blood ranged from 96.0% to 101.2%.