Dosimetric Properties of the Glass Dosimeters Used for Dose Auditing in Radiotherapy

YI Yan-ling,ZHANG Hui,CHEN Bo,ZHUO Wei-hai
DOI: https://doi.org/10.13213/j.cnki.jeom.2009.06.033
2009-01-01
Abstract:[Objective]To study the dosimetric properties of GD-301 glass dosimeter to provide a new measure for dose auditing in radiotherapy.[Methods]Five groups of the glass dosimeters were irradiated with different levels of gamma exposure from a ~(60)Co source in a secondly standard dosimetry laboratory.Based on the readout values of the irradiated glass dosimeters,the dose linearity,uniformity and reproducibility of signal as well as the fading effect of the dosimeters were analyzed.[Results] The results showed that the dose linearity of GD-301 glass dosimeter was ideal from 1 Gy to 3 Gy of irradiation.The relative deviations between the measured and reference values ranged from -0.06%to 1.9%among the 5 groups of dosimeters,and the dispersion values among 6 different dosimeters in the same groups varied from 0.4%to 1.8%.The fading effect was less than 0.9% in 30 days after irradiation.[Conclusion]Both the dose linearity and the uniformity of the GD-301 glass dosimeter are fairly good,it can be used for dose auditing in radiotherapy.
What problem does this paper attempt to address?