Value of spiral computed tomographic pulmonary angiography (CTPA) and measurement of plasma D-dimer in diagnosis of pulmonary embolism

LI Shuang-bao,TANG Hua-ping,ZHANG Ming-yong,WANG Hong-mei,HAO Yue-qin,HAO Wan-ming
DOI: https://doi.org/10.3969/j.issn.1005-8982.2007.10.028
2007-01-01
Abstract:[Object] To evaluate the clinical value of sixteen slice spiral computed tomographic pulmonary angiography (CTPA) and measurement of plasma D-dimer in the diagnosis of pulmonary embolism (PE). [Method] 104 cases suspected of having PE received CTPA, and measurement of plasma D-dimer. The diagnosis of PE is based on clinical data. [Results] 44 patients were confirmed to have PE, and 60 patients were excluded. CTPA revealed PE in 86 cases (82.17%). The sensitivity of CTPA for the diagnosis of PE is 88.6%, specificity 95.5%. The sensitivity of measurement of plasma D-dimer for the diagnosis of PE is 95.5%, specificity 63.3%. If the exclusive criteria of PE in unconfirmed diagnostic cases with CTPA is D-dimer<500 mg/L, the specificity and accuracy for diagnosis of PE increased to 88.9%, 86.4%. [Conclusions] When the unconfirmed diagnostic results with CTPA, the measurement of plasma D-dimer may be the exclusive criteria, and combined examination may increase the specificity and accuracy of diagnosis of PE.