Determination of Andrographolide and Dehydroandrographolide in Rabbit Plasma by On-Line Solid Phase Extraction of High-Performance Liquid Chromatography

Ligang Chen,Aimin Yu,Xiaodong Zhuang,Kui Zhang,Xiupin Wang,Lan Ding,Hanqi Zhang
DOI: https://doi.org/10.1016/j.talanta.2007.05.043
IF: 6.1
2007-01-01
Talanta
Abstract:The high-performance liquid chromatography (HPLC) coupled with on-line solid phase extraction (SPE) and ultraviolet (UV) detection was developed for determining andrographolide and dehydroandrographolide in rabbit plasma. Plasma samples (100μL) were injected directly into a C18 SPE column and the biological matrix was washed out for 6min using 15% aqueous methanol. By rotation of the switching valve, andrographolide and dehydroandrographolide were eluted in the back-flush mode and transferred to the analytical column by the chromatographic mobile phase consisted of methanol:acetonitrile (ACN):water (50:10:40; v/v). The UV detection was performed at 225nm. The calibration curves showed excellent linear relationship (R≥0.9993) over the concentration range of 0.05–5.0μgmL−1. The within- and between-day precisions (R.S.D.) of two analytes were in the range of 1.2–6.5% and the accuracies were between 92.0% and 102.1%. Their recoveries were all greater than 94%. The limits of detection were 0.019μgmL−1 for andrographolide and 0.022μgmL−1 for dehydroandrographolide. This method was successfully applied to the plasma concentration–time curve study after oral administration of Andrographis paniculata Nees extract in rabbit.
What problem does this paper attempt to address?