Abstract:Objective: To compare the results of detecting HBV DNA level in serum by different methods. Methods: A total of 151 Samples were selected and detected separately by COBAS Amplicor,Real-Time PCR (LightCycler and Mx3000p). Results: ① Detection of sensitivity:COBAS Amplicoris was higher than Real-Time PCR (P=0.005,0.014), while Mx3000p ,higher than LightCycler (P=0.042) ② Correlation: The results of Real-Time PCR (LightCycler and Mx3000p) and COBAS Amplicor all showed a linear correlation (r=0.842,P0.01;r= 0.946,P0.01). Conclusion: Real-Time PCR is an economical and accurate method to detect HBV DNA in the serum,and Mx3000p is superior to LightCycler in accuracy and sensitivity.