In vitro comparison of fatigue failure between conventional denture and a new type resin-bonded fixed partial denture

Jiang Ting,Lin Baoshan,Kong Ninghua
DOI: https://doi.org/10.3969/j.issn.1001-3733.2006.06.011
2006-01-01
Abstract:Objective:To compare the fatigue failure characteristics of conventional resin-bonded fixed partial dentures (C-RBFPDs) with that of a new type RBFPDs incorporating mini-anchor attachment (Crownless Bridge Work system, CBW). Methods:Ten Ni-Cr alloy conventional RBFPDs and ten Ni-Cr alloy CBW RBFPDs were respectively bonded to the acid-etched abutment teeth. All the specimens were fixed on a thermocycling and mechanical loading (TCML) chewing simulator. After thermocycling and dynamic mechanical loading (6 000× 5 ℃/55 ℃, 6×10~6×50 N, 1.7 Hz), the specimens were immersed in a 10 ml/L gentian violet solution for 24 hours,then rinsed and incised. Section surfaces were observed with an optical microscopic. Failure and crack characteristics at the metal-resin-enamel adhesion interfaces were assessed with a 4-level dye penetration microleakage scale. Statistical differences were identified with Mann-Whitney test (α = 0.05 ). Results:There were two conventional RBFPDs de-bonded completely and one with single retainer loosen from the abutment teeth. No de-bonding was found in the CBW RBFPDs group. The degree of microleakage in the conventional RBFPD group was significantly higher than that in CBW RBFPD(P=0.046). Cracks occurred mostly between the metal-resin interfaces in both groups. Conclusion:CBW RBFPDs are stronger in the resistance of fatigue failure than the conventional RBFPDs.
What problem does this paper attempt to address?