Determination of Phenol in Landfill Leachate by Using Microchip Capillary Electrophoresis with End-Channel Amperometric Detection.

YY Wu,JM Lin
DOI: https://doi.org/10.1002/jssc.200500235
IF: 3.614
2006-01-01
Journal of Separation Science
Abstract:Phenol, 2,4-dichlorophenol (2,4-DCP), and 2,4,6-trichlorophenol (2,4,6-TCP) were baseline separated by using a homemade microchip CE with an end-channel amperometric detector where a 50 microm Pt microdisk working electrode (WE) and a Pt cathode were integrated onto the microchip itself. Separation parameters such as injection time and voltage, pH of the buffer, online pretreatment condition for WE, reproducibility, and detection potential were investigated. Under the selected separation conditions, the linear ranges for phenol, 2,4-DCP, and 2,4,6-TCP were 2-200, 4-400, and 4-400 microM, respectively. The LODs were 0.4, 0.5, and 0.7 microM for phenol, 2,4-DCP, and 2,4,6-TCP, respectively (S/N = 3). The standard addition method was successfully applied to the analysis of landfill leachate samples and the concentration of phenol in the landfill leachate samples was measured to be 0.32 and 0.21 mM, respectively. The recoveries were in the range of 85-103% and corresponding RSDs were less than 5.5%.