Monitoring of Intraocular Pressure with a Sphygmomanometer during the Process of Establishment of Rat Model of Acute Intraocular Hypertension

Qin-jin CHEN,Huai-jin GUAN,Zheng-gao XIE,Jian-quan SUN
DOI: https://doi.org/10.3969/j.issn.1671-2420.2007.04.003
2007-01-01
Abstract:Purpose To investigate a method to monitor the intraocular pressure(IOP)successively during the process of establishment of rat model of acute intraocular hypertension.Methods The rat model of acute intraocular hy- pertension was established through perfusion of balance salt solution into the anterior chamber of the rat eye.The IOP was measured with a sphygmomanometer that was connected with a plastic tube inserted into the anterior chamber.The differ- ence between IOP when the ambient was connected with the solution surface directly or indirectly and the formed pressure by the perfused solution was observed.Results There was no significant difference between the IOP of rats and the pres- sures the perfused solution should produce(P>0.05)when these super-surfaces of balanced salt solution were exposed to the air directly;but there were significant difference(P<0.05)when these super-surfaces of balanced salt solution were exposed to the air indirectly.Conclusion The IOP can be monitored accurately and successively by a sphygmoma- nometer during the process of establishment of a model of acute intraocular hypertension.The vacuum in the bottle may influence the IOP.(Chin J Ophthalmol and Otorhinolaryngo1,2007,7:211-213)
What problem does this paper attempt to address?