Rapid Determination of Rifampicin and Isoniazide Tablets Using Near Infrared Diffuse Reflectance Spectroscopy

LU Jia-hui,CHEN Yu-fei,JIANG Chao-jun,SHI Jian,GUO Wei-liang,TENG Li-rong
DOI: https://doi.org/10.3321/j.issn:1671-5489.2007.03.037
2007-01-01
Abstract:Objective:To establish a rapid method to determination of rifampicin and isoniazide in rifampicin and isoniazide tablets by near infrared technology.Methods:NIR spectra were pretreated by Savitzky-Golay,1 st order derivative and 2nd order derivative respectively;and then were used to construct quantitative analysis models using PLS.Results:In the calibration models,the regression coefficient(r)in rifampicin determination was 0.9940,the root mean square error of cross-calibration(RMSECV)was 0.00764,the r in isoniazide determination was 0.9910,and the RMSECV was 0.00445.The samples of validation set were predicted by these models.The root mean square error of prediction(RMSEP)in rifampicin determination was 0.00397.The RMSEP in isoniazide de- termination was 0.00383.This result demonstrates that the method has high precision of prediction.The average re- coveries of rifampiein and isoniazide are 100.88% and 100.87% respectively,and the RSDs were 0.672% , 1.08% in reappearance experiments.Conclusions:With the advantages of facility,rapid,non-destruction,no pol- lution,on-line detection and good reappearance,the method must be popularized in situ measurement and in the on-line quality control for rifampicin and isoniazide tablets productions.
What problem does this paper attempt to address?