Study on Methodology of Fingerprint of Gas Charomatography of Liquor

Tian-xiang WU,Yan ZHENG,Qing-li TANG
2008-01-01
Abstract:The methods and conditions for the establishment of fingerprint of GC of liquor were studied and the results showed that the optimum conditions were as follows:flow velocity of carrier gas (notrogen) was 1.3 mL/min and the split ratio was 50∶1. Meanwhile, the established methodology was investigated and RSD of the relative retention time for each chromatographic peak was less than 0.03 % and RSD of relative peak area was less than 5.0 %. The stability test of samples indicated that RSD of peak retention time for each chromatographic peak was less than 0.04 % and RSD of relative peak area was less than 5.0 %. The precision of the apparatus, reliable experimental methods, and stable samples were the basic technical requirements for fingerprint.