Electromagnetic Navigation-Guided Acupuncture of Foramen Ovale: A Cadaver Study

陈敏洁,杨驰,顾力栩,张伟杰,董敏俊,赵晶,姜滨,邱亚汀,冯志强
DOI: https://doi.org/10.3969/j.issn.1001-3733.2010.01.012
2010-01-01
Abstract:Objective: To evaluate the surgView-RFT electromagnetic navigation system in radiofrequency thermocoagulation(RFT).Methods: 6 foramen ovales of 3 cadavers were punctured by 3 beginners.Every one tried 5 times by free-hand and 5 times by SurgView-RFT electromagnetic navigation-guidance.After each puncture,CT scans were used as golden standard to measure the distance between the tip of needle and the anterior margin of the foramen ovale.T test and variance of SAS 6.12 statistical software was used to do the statistic analysis.Results: The punctures by free-hand failed,and the distances between the tip of needle and the anterior margin of the foramen ovale were 8-10.6 mm.The punctures by navigation guidance were successful,and the distances between the tip of needle and the anterior margin of the foramen ovale were 2.68-3.54 mm.The error of navigation system was 0.59 mm which was verified by CT scan.There was significant difference between free-hand puncture and navigation-guided puncture.Conclusion: The surgView-RFT electromagnetic navigation system is characteristic with high performance,high accuracy and minimal invasion,while it is applied in training or clinical using.