Influence of Soft Tissue Shift in Electromagnetic Navigation-guided Acupuncture of Skull Base: an Animal Study

陈敏洁,杨驰,顾力栩,张伟杰,董敏俊,赵晶,张晓虎,邱亚汀,冯志强
2009-01-01
Abstract:Objective: To evaluate the influence of buccal soft tissue shift on the electromagnetic navigation-guided acupuncture of skull base.Methods: 6 inferior orbital fissures(IOF) of 3 living goats were punctured by 3 beginners.Every one tried by free-hand and by electromagnetic navigation-guided.After each puncture,CT scans were used to measure the distance between the tip of needle and the center of the IOF.t test and variance of SAS 6.12 statistical software was used to analyse. Results: The punctures were failed by free-hand,and the average distance between the tip of needle and the center of the IOF was 15mm in the left sides and 14.3mm in right sides.The punctures were successful by navigation guide,and the average distance between the tip of needle and the center of the IOF was 2.7 mm in the left sides and 2.8 mm in right sides.The average range deviation and the average angular deviation in the buccal tissue were 4.3 mm and 6.2mm.There was statistical difference between free-hand puncture and navigation-guided puncture(P range from 0.0008 to 0.0239).Conclusion: The surgView-RFT electromagnetic navigation system can overcome the deviations of buccal soft tissue shift,due to the target was bony structure.It could be used in radiofrequency thermocoagulation,fine needle biopsy of tumor at skull base or drainage of abscess in deep facial part with minimal error.