Reproducibility of Pulse Wave Measurement

YAN Liu-xia,LI Ying,LI Xian,ZHAO Lian-cheng,WU Yang-feng
DOI: https://doi.org/10.3969/j.issn.1673-7245.2007.04.017
2007-01-01
Abstract:Objective To assess the reliability of central and peripheral pulse wave velocity measurement. Methods Carotid-femoral and carotid-radial pulse wave velocity(CFPWV, CRPWV) measurements were performed in five healthy female volunteers by two observers in two separate days, with 3 exams each day at 9:00, 11:00, 14:00 respectively. Results The differences of mean absolute values between two observers were 0.07±0.70 m/s(9.54±1.02 m/s vs 9.47±0.82 m/s) and 0.64±1.40 m/s(9.60±1.10 m/s vs 8.96±0.91 m/s) for the CFPWV and CRPWV. Pearson coefficients were 0.73(P<0.01) and 0.1(P=0.86) and total intra-class coefficients were 0.50 and 0.15 for CFPWV and CRPWV. Bland-Altman plots showed good inter-observer reproducibility both for CFPWV and CRPWV. The inter-occasion reproducibility was not content enough for clinical application, with intra-class coefficient 0.23-0.68 and 0.01-0.32 for CFPWV and CRPWV. Conclusion Under standardized measurement conditions, CFPWV and CRPWV showed good inter-observer reproducibility but less satisfactory inter-occasion reproducibility.
What problem does this paper attempt to address?