Research on the quantitative determination of main ingredients in soybean and maize by Fabry-Perot interferometer NIR spectrometer

Wang Dong,Min Shungeng,Zhu Yewei,Wu Lijun,Li Gaoqian,Tang Guo,Li Chunzi
DOI: https://doi.org/10.3969/j.issn.1672-7916.2011.05.008
2011-01-01
Abstract:The NIR spectra of 40 soybean samples and 46 maize samples were collected by Axsun portable NIR spectrometer.The quantitative calibration models of crude protein and fat of soybean and the quantitative calibration models of crude protein,fat and starch of maize were developed by PLS algorithm respectively.The result indicated that for the quantitative calibration models of soybean crude protein and soybean fat,the model dimensions are 6 and 5 respectively,the determination coefficients (R2) are 97.94% and 93.66% respectively, the root mean square errors of calibration (RMSEC) are 0.40 and 0.36 respectively,the root mean square errors of cross validation (RMSECV) are 0.49 and 0.44 respectively,the relative performance deviations (RPD) are 5.18 and 3.15 respectively;for the quantitative calibration models of maize crude protein,maize fat and maize starch,the model dimensions are 4,4 and 6 respectively,R2 are 90.15%,95.22% and 87.81% respectively,RMSEC are 0.25,0.12 and 0.53 respectively,RMSECV are 0.37,0.15,0.72 respectively,RPD are 2.57,3.57 and 2.42 respectively.The result of F-test indicated that a very remarkable correlation exists between the estimated and specified values of each calibration model mentioned above.This research indicated that the spectrometer can be applied in the on-site rapid quantitative determination of the main ingredients of soybean and maize.
What problem does this paper attempt to address?