Development of high-performance magnetic chemiluminescence enzyme immunoassay for α-fetoprotein (AFP) in human serum

Xu Wang,Qian-Yun Zhang,Zhen-Jia Li,Xi-Tang Ying,Jin-Ming Lin
DOI: https://doi.org/10.1016/j.cca.2008.03.010
IF: 6.314
2008-01-01
Clinica Chimica Acta
Abstract:Results The proposed method had a sensitivity of 3.0 ng/ml, low cross reactivities, and an assay time of 1 h. The linear range was 0–1200 ng/ml through using MPs and is useful for samples with extremely high AFP concentrations without dilution while avoiding the hook effect. The intra- and inter-assay precision was < 3% and < 5%. The present method has been successfully applied to the detection of AFP human serum with recoveries from 90 to 108%, and showed a good correlation with the commercially available AFP RIA kit. Conclusions This proposed assay provided apparent advantages over microplate CLEIA and RIA, and facilitated the development of high-throughput screening and automated operation systems in the clinical practice. Keywords Chemiluminescence enzyme immunoassay (CLEIA) Magnetic particles (MPs) Hepatocellular carcinoma (HCC) α-Fetoprotein (AFP)