Selection of Regions for Noise Measurement in Thoraco-abdominal MDCT Angiography

YU Hong,LI Hui-min,YUAN Zheng,LI Xiao-lin
DOI: https://doi.org/10.3969/j.issn.1006-5741.2009.02.020
2009-01-01
Abstract:Purpose:The aim is to find out the appropriate region to accurately and conveniently reflect the noise of the initial images for thoraco-abdominal MDCT angiography.Methods:Forty-six consecutive patients(male 33,female 13,mean age 56.4 years,median age 55.0 years)underwent 16-MDCT(Toshiba Aquilion 16)with the scanning parameters included 120kVp,a 0.5-second gantry rotation time,a pitch of 0.938:1,and 16×1mm detector collimation.The mA was set as 150 for thorax scanning(20 patients)and 200 for abdomen scanning(26 patients).A long water phantom was placed affront of thorax or abdomen.The arterial phase images were retrospectively reconstructed with 1mm slice thickness,0.8mm interval,and FC 2.The standard deviation(SD)of the aorta at the level of right pulmonary trunk(thorax)or posterior crura of diaphragm(abdomen)was recorded.At the same section,the SD of phantom,muscle of back,and abdominal water region were also recorded.Results:The SD of the phantom was referred as standard,the SDs in aorta,muscle,and abdominal water region were significantly different(P=0.000).When a constant was subtracted,the SDs in aorta,muscle,and abdominal water region were not significantly different to that in phantom(P>0.05).Conclusion:The aorta,muscle,and water region were all adaptable to predict the image noise,but a modification was needed for accurate estimation.The aorta would be the best one and were recommended.