Preparation of QD-CdS/TiO2 photoelectrochemical sensor and its application in the detection of chemical oxygen demand

Yan-ju Fang,Hong-chun Ding,Zhong-hai Zhang,Zhen-sheng Pan,Li-tong Jin
DOI: https://doi.org/10.3969/j.issn.1008-2298.2005.04.003
2005-01-01
Chemical Sensors
Abstract:In this work,a QD-CdS/TiO_2 photoelectrochemical sensor was prepared and applied to chemical oxygen demand(COD)determination based on the charge transfer on the interface of QD-CdS/TiO_2 film electrode and the passing solution.With a positive bias potential of 0.4V(vs.Ag/AgCl)applied to the sensor and an 11-W quartz UV lamp illuminating it,photocurrent,was recorded and its change(△I_(photo))caused by the detected sample was calculated to characterize the COD value of the sample.Under the optimizing conditions,the sensor responded linearly to the COD of D-glucose solution in the range of 15.0~600.0mg/L and the lowest limit of detection is 6.5mg/L.