Self-Organized Device for Photoelectrocatalytic Determination of Chemical Oxygen Demand and Performance Research

Yin Weiping,Li Xuejin,Zhou Baoxue
2013-01-01
Abstract:A set of automatic data acquisition and processing device was assembled for water COD value determination,which is based on the photocatalytic reaction principle and the TiO 2 nanotube array electrode.The results of the research for the influencing factors showed that the device can get stable results with 365nm ultraviolet light source,2.0V bias voltage,2.0mol/L NaNO 3 solution as supporting electrolyte,in the pH range from 4 to 10.COD measurement range is from 0 to 500mg/L,the detection limit is 0.3mg/L,in the range,the net charges and the theoretical COD are in good linear relationship,the equation of which is Q net =0.01308ThCOD+0.02989,the correlation coefficient is 0.9962.COD values obtained by this new method are compared with the standard potassium dichromate method.The relative deviations of different kinds of samples are both less than 4%.So,the standard way of potassium dichromate method can be replaced by the new one to overcome the shortcomings.
What problem does this paper attempt to address?