Comparison of CD4T cell counts measured by flow cytometry of BD FACS Calibur and Beckman Coulter XL

彭虹,徐建青,段丹丽,袁霖,绳波,王露,阮玉华,韩丽凤,苏斌,徐臣,邢辉,邵一鸣
DOI: https://doi.org/10.3969/j.issn.1672-5662.2006.03.003
2006-01-01
Abstract:Objective Since flow cytometry has been widely used in the national effort to contain AIDS epidemic in China,it is important and urgent to understand and compare CD_4 T-cell counts determined by different types of this apparatus.Methods To gain information in this aspect,we used two types of flow cytometry,i.e.,BD FACS Calibur and Beckman Coulter Xl in parallel,to quantify CD_4T-cell counts in 183 blood samples from Anhui province,including 41 samples from seronegative individuals and 142 samples from HIV-infected subjects.Result Overall,the difference in 57% of samples from 41 seronegative individuals was<10% while in 43% of samples it was>10%.In contrast,119 samples out of 142 ones(84%)in HIV-infected individuals generated<10% difference by using these two types of flow cytometry,whereas only 16% of samples produced >10% discrepancy.The correlation coefficient was equal to 0.97 in the seronegative group and 0.98 in the HIV-infected group.Statistical analysis showed no significant difference between data generated by these two types of flow cytometry(P>0.05).Conclusion Data generated by these two types of flow cytometry are highly correlated and can be directly used at a range of CD_4T-cell counts between 50 and 500.