Analysis of Ptosis in Patients with Myasthenia Gravis Using a Video Eye-Measuring System

HAO Hongjun,JIAO Hongyun,ZHAO Guiping,GAO Feng,HUANG Yining
DOI: https://doi.org/10.13505/j.1007-1482.2010.02.013
2010-01-01
Abstract:Objective In order to provide a direct, accurate and standard measuring method for clinical diagnosis of myasthenia gravis (MG,), we recorded the changes of eyelid drooping degree after the injection of neostigmine with a video eye-measuring system. Methods Thirty-four patients with MG, 30 cases with non-MG disease and 30 normal helthy subjects were included in this study. The palpebral fissure size of patients in the MG group, non-MG control group and normal control group before and after the injection of neostigmine, analyzing the repeated value and effectively relative value, and then calculated the sensitivity, specificity and Youden index of the repeated values. Results The mean ± standard deviation value of palpebral fissure size of orthophoria was 2.621±0.376 mm, 0.823±0.266 mm and 0.853±0.254 mm in patients of the MG, non-MG control and normal control groups, respectively. There was a significant difference (P0.05) among patients of the MG, non-MG control and normal control groups. The sensitivity, specificity and Youden index of the repeated values was 96.7%, 100% and 0.967, respectively. Conclusions The application of video eye-measuring system with the neostigmine test is a direct ,accurate and standard method for the diagnosis of MG.
What problem does this paper attempt to address?