Determination of PAH by Solid-phase Microextraction Coupled to HPLC Using Fibers with Mesoporous Coating

LIU Yan,YANG Fu-wei,YANG Ling-juan
DOI: https://doi.org/10.3969/j.issn.1005-8141.2011.07.004
2011-01-01
Abstract:The solid-phase microextraction(SPME) using C8-SBA-15 fiber coating coupling with high performance liquid chromatography(HPLC) was applied to determine trace of environment water.The experimental conditions of SPME,such as extraction time,extraction temperature,desorption time were optimized.The optimized conditions were as below: room temperature,400r/min of stirring rate,30 min of extraction time and 3min of desorption time,respectively.The linear range was 0.4—300μg/L,4.0—450μg/L.The limit of detection was 0.3μg/L,0.10μg/L.The sampling procedures were simple,rapid and solvent free,and the technique had high sensitivity and good precise.
What problem does this paper attempt to address?