Gold nanoparticle-attached ITO as a biocompatible matrix for myoglobin immobilization: direct electrochemistry and catalysis to hydrogen peroxide

Jingdong Zhang,Munetaka Oyama
DOI: https://doi.org/10.1016/j.jelechem.2004.12.005
IF: 4.598
2005-01-01
Journal of Electroanalytical Chemistry
Abstract:A gold nanoparticle-attached indium tin oxide (Au/ITO) electrode was applied for myoglobin (Mb) immobilization. The direct electron transfer of Mb was shown by the stable and well-behaved voltammetric responses for Mb immobilized on the Au/ITO electrode in buffer solutions. The effects of scan rate, pH value and oxygen on the direct electrochemistry of Mb were observed. Based on the catalytic activity of Mb immobilized on a gold nanoparticle-modified ITO (Mb/Au/ITO) electrode toward the reduction of H2O2, a mediator-free H2O2 sensor was developed, which showed good reproducibility and stability in pH 7.0 buffer. A linear relationship existed between the catalytic current and the H2O2 concentration in the range of 2.5×10−6–5×10−4 M with a detection limit (S/N=3) of 4.8×10−7 M.