146 Improved 3D SPACE carotid vessel wall imaging at 3.0 T

Zhaoyang Fan,Zhuoli Zhang,Yiucho Chung,Peter Weale,Ioannis Koktzoglou,Sven Zuehlsdorff,Qi Yang,Kuncheng Li,John Sheehan,Timothy Carroll,Jin An,Xun Zhang,Qiang Zhang,Renate Jerecic,James Carr,Debiao Li
DOI: https://doi.org/10.1186/1532-429X-10-S1-A47
IF: 6.4
2008-01-01
Journal of Cardiovascular Magnetic Resonance
Abstract:Introduction Suppression of the intraluminal blood signal is essential for visualizing arterial wall. 2D turbo spin-echo (TSE) acquisition coupled with a double inversion-recovery (DIR) preparation has been the method of choice for arterial wall imaging. 2D techniques, however, have relatively low through-plane resolution and limited coverage as compared to their 3D counterpart. Two challenges of 3D techniques include blood signal suppression and long imaging time. Recently, a motion-sensitive dephasing (MSD) preparation scheme was used for blood suppression in conjunction with 3D steady-state free precession (SSFP) acquisition [1]. However, at 3.0 T, SSFP suffers from off-resonance artifacts. Variable-flip-angle TSE (SPACE) has been introduced as an alternative for fast imaging of carotid arterial wall at 3.0 T due to its insensitivity to field inhomogeneities [2]. Although most intraluminal signal is suppressed due to its inherent flow spin dephasing effect, residual blood signal is often observed in regions of slow or complex flow, particularly at the carotid bifurcation. We proposed to use MSD preparation in SPACE to improve its blood suppression efficiency.