Simultaneous T 1 and T 2 mapping of the carotid plaque (SIMPLE) with T 2 and inversion recovery prepared 3D radial imaging.

Haikun Qi,Jie Sun,Huiyu Qiao,Xihai Zhao,Rui Guo,Niranjan Balu,Chun Yuan,Huijun Chen
DOI: https://doi.org/10.1002/mrm.27361
2018-01-01
Magnetic Resonance in Medicine
Abstract:Purpose: To propose a technique that can produce different T-1 and T-2 contrasts in a single scan for simultaneous T-1 and T-2 mapping of the carotid plaque (SIMPLE). Methods: An interleaved 3D golden angle radial trajectory was used in conjunction with T-2 preparation with variable duration (TEprep) and inversion recovery pulses. Sliding window reconstruction was adopted to reconstruct images at different inversion delay time and TEprep for joint T-1 and T-2 fitting. In the fitting procedure, a rapid B-1 correction method was presented. The accuracy of SIMPLE was investigated in phantom experiments. In vivo scans were performed on 5 healthy volunteers with 2 scans each, and on 5 patients with carotid atherosclerosis. Results: The phantom T-1 and T-2 estimations of SIMPLE agreed well with the standard methods with the percentage difference smaller than 7.1%. In vivo T-1 and T-2 for normal carotid vessel wall were 1213 +/- 48.3 ms and 51.1 +/- 1.7 ms, with good interscan repeatability. Alternations of T-1 and T-2 in plaque regions were in agreement with the conventional multicontrast imaging findings. Conclusion: The proposed SIMPLE allows simultaneous T-1 and T-2 mapping of the carotid artery in less than 10 minutes, serving as a quantitative tool with good accuracy and reproducibility for plaque characterization.