High-performance Liquid Chromatography Study of Gatifloxacin and Sparfloxacin Using Erythrosine As Post-Column Resonance Rayleigh Scattering Reagent and Mechanism Study.
Ziyu Pan,Jingdong Peng,Xu Zang,Huanjun Peng,Huan Xiao,Lingli Bu,Fang Chen,Yan He,Yu Chen,Xiang Wang,Shiyu Li,Yi Chen
DOI: https://doi.org/10.1002/bio.3429
IF: 2.613
2017-01-01
Luminescence
Abstract:Herein, a highly selective high-performance liquid chromatography (HPLC) coupled with resonance Rayleigh scattering (RRS) method was developed to detect gatifloxacin (GFLX) and sparfloxacin (SPLX). GFLX and SPLX were first separated by HPLC, then, in pH4.4 Britton-Robinson (BR) buffer medium, protonic quaternary ammonia cation of GFLX and SPLX reacted with erythrosine (ERY) to form 1:1 ion-association complexes, which resulted in a significant enhancement of RRS signal. The experimental conditions of HPLC and post-column RRS have been investigated, including detection wavelength, flow rate, pH, reacting tube length and reaction temperature. Reaction mechanism were studied in detail by calculating the distribution fraction. The maximum RRS signals for GFLX and SPLX were recorded at (ex)=(em)=330nm. The detection limits were 3.8ng ml(-1) for GFLX and 17.5ng ml(-1) for SPLX at a signal-to-noise ratio of 3. The developed method was successfully applied to the determination of GFLX and SPLX in water samples. Recoveries from spiked water samples were 97.56-98.85%.