Elemental mass spectrometry using a helium microwave plasma torch as an ion source

Min Wu,Yixiang Duan,Qinhan Jin,Gary M. Hieftje
DOI: https://doi.org/10.1016/0584-8547(94)80013-8
1994-01-01
Abstract:An investigation was performed to evaluate the coupling of a 150 W atmospheric pressure helium microwave plasma torch (MPT) with mass spectrometric detection. The interface of an existing inductively coupled plasma mass spectrometer was modified to sample the helium plasma. Optimization studies illustrate the dependence of analytical signals on various instrumental parameters, including sampling depth, gas and sample solution flow rates, and ion optic settings. With the use of a pneumatic nebulization-desolvation sample introduction system, detection limits for seven non-metallic elements range from 12 ng/ml to 1.0 mug/ml, comparable to those determined with an Ar ICP and the same mass spectrometer. Interference effects of concomitant cations and isotope ratios for Cl and Br are also presented.