Determination of High Molecular Mass Impurities in Cefuroxime Axetil and It's Preparation by Polystyrene- Based Gel Filtration Chromatography

HOU Yu-rong,YUAN Yao-zuo,ZHANG Mei,QIAN Wen,HANG Tai-jun
DOI: https://doi.org/10.16155/j.0254-1793.2012.02.028
2012-01-01
Abstract:Objective:To determinate the high molecular mass impurities in cefuroxime axetil by polystyrene-based gel filtration chromatography.Method:A MKF-GPC-100(7.8 mm×100 mm,8 μm) column was adopt with the mobile phase consisting of water-methanol(10∶ 90),at a flow rate of 0.5 mL·min-1,and the detection wavelength was at 278 nm,to identify the impurities by HPLC-ESI-MS2.Result:The calibration curves for cefuroxime axetil in the range of 0.5-1500 μg·mL-1 was linear(r=1.000),the detection limit was 0.2 μg·mL-1;the peaks before cefuroxime axetil in the chromatogram were high molecular mass impurities,the resolution between cefuroxime axetil and high molecular mass impurities was good.The high molecular mass impurities is instable in solution,and the test solution should be prepared immediately before use.Conclusion:The established method is fast for analysis and good for precision,it's suitable for determinate the high molecular mass impurities in liposoluble drug cefuroxime axetil and it's preparation.