Study on Responding Time of α Rapid Radioactive Aerosol Monitors

ZHANG Yan,LI Yang,XUE Tao
DOI: https://doi.org/10.3969/j.issn.0258-0934.2013.04.002
2013-01-01
Abstract:Energy discrimination is an important method to eliminate the interference of measurement caused by natural radioactivity. Through the methods of theoretical derivation as well as numerical calculation,the article established the relationships between responding time of radioactive aerosol monitor with equalization coefficient K,instrument detecting efficiency,sampling flow,detection limit as well as the concentration of radon daughter. Results show that responding time of radioactive aerosol monitor can be shorten by decreasing equalization coefficient K,improving instrument detecting efficiency,increasing sampling flow,raising detection limit of artificial radioactive aerosol and lowering concentration of natural radioactive aerosol. This is beneficial to optimizing the design of radioactive aerosol monitor.
What problem does this paper attempt to address?