Determination of ascorbic acid in drug formulations using sequential injection analysis with chemiluminescence detection

XU Xiao-na,WANG Nai-zhi,FAN Shi-hua
DOI: https://doi.org/10.3969/j.issn.1000-0720.2006.01.020
2006-01-01
Abstract:A sequential-injection chemiluminescence method was applied to the assay of ascorbic acid in drug formulations.The method is based on the chemiluminescence reaction of ascorbic acid with KMnO_4 in HCl media,using HCHO as an enhancer.The linear range was from 1.0×10~(-9) to 2.0×10~(-6) mol/L with the detection limit of 5.0×10~(-10) mol/L(3σ).The correlation coefficient of 0.9995(n=5) and relative standard deviation(R.S.D) of 1.3%(n=11) were achieved for 1.0×10~(-7) mol/L ascorbic acid.The method was successfully used to determine the content of ascorbic acid in tablets and injections and the results were consistent with those of the standard method.
What problem does this paper attempt to address?